Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 1 de 1
Filtrar
Mais filtros










Base de dados
Intervalo de ano de publicação
1.
Health Phys ; 112(3): 237-245, 2017 03.
Artigo em Inglês | MEDLINE | ID: mdl-28121723

RESUMO

Optically-stimulated luminescent dosimeters are capable of being interrogated multiple times post-irradiation. Each interrogation removes a fraction of the signal stored within the optically-stimulated luminescent dosimeter. This signal loss must be corrected to avoid systematic errors in estimating the average signal of a series of optically-stimulated luminescent dosimeter interrogations and requires a minimum number of consecutive readings to determine an average signal that is within a desired accuracy of the true signal with a desired statistical confidence. This paper establishes a technical basis for determining the required number of readings for a particular application of these dosimeters when using certain OSL dosimetry systems.


Assuntos
Artefatos , Guias como Assunto , Iluminação/normas , Dispositivos Ópticos/normas , Dosimetria Termoluminescente/instrumentação , Dosimetria Termoluminescente/normas , Calibragem/normas , Desenho de Equipamento , Análise de Falha de Equipamento , Iluminação/instrumentação , Doses de Radiação , Reprodutibilidade dos Testes , Semicondutores , Sensibilidade e Especificidade
SELEÇÃO DE REFERÊNCIAS
DETALHE DA PESQUISA
...